Surgical Cautery (CORE- 250 - watt) |
|
| |
|
|
|
| |
|
High Performance For Most Demanding Procedures With Quick Response Technology
In The Operation Theater, Clinic & Outpatient Surgery Center The Professional Core – 250 Electro Surgical Generator Provides A High Level Of Performance And Safety. The Core – 250 Accepts All Standard Accessories To Help Meet The Surgeon’s Needs In Wide Range of Demanding Procedures Like Laparoscopy , Urology , General Surgery , Orthopedics , Gynecology Neuro- Surgery Pediatric , Plastic Cardio Thoracic And Reconstructive Surgeries.
|
|
 |
|
|
| |
|
Advanced Performance Features :- |
| |
|
Monopolor Quick Response Technology |
| |
|
|
| |
|
(A)
Two Cut Mode |
|
| |
|
- Pure cut [With Repaid Start Cutting] Suitable For Under Water Cutting (TURP , TUVPS , TCRE) With uro- cut Which Elilminates Lag Time In Initiating Cut Especially Suited For Urological Procedures.
- Blend cut Preset Or With Variable Blend Option For Cutting With Varying Degree Of Homoeostasis SUITED FOR LLETZ / LEEP Gyne Procedures.
|
|
| |
|
(B) Two Coag Modes |
|
| |
|
- Spray Coag For Non Contact Large Tissue Area Coagulation With Minimum Depth Of Necrosis.
- Desiccation Coag For Low Voltage Soft Coagulation Suitable For Laproscopic And Delicate Tissue Work
|
|
| |
|
Bipolar – SlS (Self Limiting System – Unmodulated) |
|
| |
|
- Micro For Delicate Tissue Work By Controlled Desiccation Without Charring, Adhesion Or Blanching Of Adjacent Tissue [Opthalmic & Neuro - Surgery].
- Macro For New Generation Of Macro Bipolar Cutting Instruments [Tri - Polar] For Lavh Procedurc. With High Pressure Vessel System [VLS].
|
|
| |
|
|
|
| |
|
Advanced Safety Features:- |
|
| |
|
|
|
| |
|
- Isolated RF Output Direct Current Return Back to The Generator to Reduce The Risk of Alternate Side Burnd and to Ensure Patients Safety In Case of any Discontinuity In Patients Plate Cord.
- PEM- Patient Electrode Monitoring [Optional]
- Class ‘2’ Leakage Cancelling A Must for Lengthy Procedure Like Cardiac Surgery [CABG] etc.
- Independent Activation of Hand Switching. Foot Switching & Bipolar Output to Protect Against In Advertent Output From an Idle Accessory Lying on or Near The Patient or Surgical Staff.
|
|
| |
|
Enhanced Operating Convenience :-
-
Two Independence Hand Swithching Mono Polar Outputs And One Hand Switching Bipolar Output Combined With Lapro-Scopic Clipons Elements The Need Of Foot Switching Durring Endoscopic Procedures.
-
Separate Bipolar Generator Controls And Foot Switch With Auto Selectivity Reduces The Dependency On The Operating Room Staff During The Procedure.
-
Dual Generator System – Two Independent Mono Polar Hand Switching Outputs.
-
Ergonomically Designed Trolley To House Diathermy , Accessories , Argon Unit With Cylinders And Smoke Evacuator.[Optional]
-
Comprehensive Clinical Training Network Through Workshops, Seminars, Videos & Literatures.
-
Smart Modular Technology For Last & Effective After Sales Supports (Service)
|
|
 |
|
|
| |
|
|
|
| |
|
Technical Specification :- CORE 250 (220 -240) |
|
| |
|
|
|
| |
|
Weight and Dimensions |
|
| |
|
Height 7 Inch [17 cm] |
|
| |
|
Width 13.5 Inch [32cm] |
|
| |
|
Length 16.5 Inch [40cm] |
|
| |
|
Weight 9.5 Kg |
|
| |
|
|
|
| |
|
Low Frequency Leakage :- Patient Leakage, 150 mA rms |
|
| |
|
Cooling :- Single Speed Fan |
|
| |
|
High Frequency Leakage :- LESS THAN 150 mA rms, All Output |
|
| |
|
Input Power Requirements :- Volts: 180 -270 V AC |
|
| |
|
Amps: 6A |
|
| |
|
Input : 1.5 KW, 50/60 Hz |
|
| |
|
|
|
| |
|
Audio Volume :- |
|
| |
|
|
|
| |
|
The Mode Indicators Tones Are Adjustable To A Minimum Level of 45 dba at 1 Meter Designed To Meet UL, CSA AND IEC – 60601 -2-2 Conforms to IS – 7583s. |
|
| |
|
|
|
| |
|
Standard |
|
| |
|
|
|
| |
|
- P- 001 Hand Switching Pencil
- P- 002 Patient Plate With Cord
- P-003 Mono Polar Foot Switch
- P-004 Bipolar Foot Switch
- P-005 Chuck Handle
|
|
| |
|
|
|
| |
|
Optional |
|
| |
|
|
|
| |
|
- P-006 Bipolar Forceps with Cord
- P-007 Hand Switching Bipolar Forceps with Cord
- P-008 Mono Polar Forceps with Cord
- P-009 LLETZ / LEEP Electodes Needle, Ball Loop
- P-010 Pediatric Patients Plate With Cord
- P-011 Ligation Sure Bipolar Artery Forceps
- P-012 Lap-Hand Switching Electrodes
- P-013 Coated Vaginal Speculums With Smoke Evacuation Port
- P-014 Mobile Trolley
- P-015 HF Mono Polar Cord
- P-016 HF Bipolar Cord
- P-017 Equipotential Lug
- P-018 Smoke Evacuation Unit
- P-019 Lap Clipon Bipolar
- P-020 Lap Clipon Mono Polar
- P-021 BCM ( Bipolar Current Monitor) Module
- P-022 Triple Padle Foot Switch
|
|
| |
|
Output Characteristics :- |
|
| |
|
|
|
| |
|
| |
Mode |
Output
Fred.
(kHZ) |
Maximum
P-P Voltage |
Rated Load [Ohms] |
Max Power [Watts] |
Creast Factor |
Biopolar |
Micro Macro |
500
500 |
500
900 |
100
100 |
90
100 |
2.5
1.5 |
Monopolar
Cut |
Pure
Bleand |
500
500 |
2800
2500 |
500
500 |
400
250 |
1.5
2.4 |
Mono Polar
Coag |
Desiccate
Spray |
450
450 |
4000
8000 |
500
500 |
150
150 |
6.4
8.0 |
|
|
| |
|
|
|
| |
|
Crest Factor Is An Indicator Of A Wave Farm Ability To Coagulate The Bleeders Without Cutting Effects. |
|
| |
|
|
|
| |
|
|
|