PX- 300 & 500 |
|
| |
|
|
| |
PX - 300/500 X-Ray Unit Can Be Used For Routine Radiographic and Fluoroscopic Diagnostic Procedures of the Chest, Extremities, Skull, Skeletal System, Respiratory System, Obstetric, Pelvis, Spinal Column And Gastro Intestinal Track And For Special Investigation Including Barium, IVP And HSG Examinations.
Sailent Features :-
|
|
|
| |
- Radiographic Output: 24 KW (PX-300) and 40KW (PX-500) as per IS7620 (Part 2)
- Tube Filament Current Stabilisation Through Constant Voltage Transformer (CVT)
- Full Wave Rectification
- Digital Display of KV, mA and mAS
- Overload Protection (Blockage Of Exposure)
- Shortest Exposure Time 10 ms
- Supports Serialography upto 6fpts
- kv, mA and S valves Set With Step Switches
- Floroscopy Time Totaliser
- 5 Led Indicators for Line, O\ L ,Exposure, Ready and Bucky In,
- Voltage Stabilisation (Compensation) Through Course and Fine Voltage Compensation Switches.
- Volt Meter and mA meter for Viewing Input Line Voltage and mA during Exposure.
|
|
| |
|
|
| |
Added Applications :- |
|
| |
|
|
| |
Providing You Complete Solutions For Every Application Wide Range Of Optional High Quality Accessories Are Also Available.All With The Guarantee Of Quality And Enduring Value That Only" PROFESSIONAL" Can Provide.
|
|
| |
- Radiography and Radiography \ Fluoroscopy Floor-To- Ceiling Column Stands
- Horizontal and Manual Multi Position Bucky Examination Tables
- Motor Operated Table Spot Film Device
- Image Intensifier Television Option
- Option of Single or Double Tube Operation
|
|
| |
Technical Data For PX – 300 \ 500 :- |
|
| |
|
|
| |
Specification |
PX-300 |
PX-500 |
Output Power
(As Per IS : 7620) |
24KW |
40 KW |
Radiography
Parameters |
kV Range 40-100 kV (2 kV Each Step)
mA Range Small Focus (50 ,100)
Large Focus (200 , 300) |
kV Range 40-100 kV (2 kV EACH STEP)
mA Range Small Focus (50 ,100 ,200)
Large Focus (300 , 400 ,500) |
| Exposure Time |
0.01 to 5 Sec. |
0.01 to 5 Sec. |
Fluoro Scopy
Parameters |
kV RANGE 40 TO 90 kV in Steps mA RANGE 0 -3 mA |
| Exposure Release |
From Control Console With Two Step Switch or From Examination Unit |
| Fluorscopy Release |
From Control Console or From Examination Unit |
| Mainsvoltage Compensation |
Provided (354 -447 Volts) |
| Filament Stabilisation |
Provided |
| Over Load Protection |
Provided , Blocking For Any Hazardous Combination |
| Tube Head |
THX-1 BEL\Imported Rotating Anode Dual FOCUS 20\40 kW
OR
Dual Focus 30\50 kW (Optional) |
| HV Cable |
Pair of HV cable |
| Power Supply 3 Phase |
440V+10% - 15% 50Hz 63A |
| Main Resistance |
< . 36 OHM |
|
|
| |
|
|
|