Mfr. of Medical Equipment
 
   
  PX- 300 & 500  
 

 

 
  PX - 300/500 X-Ray Unit Can Be Used For Routine Radiographic and Fluoroscopic Diagnostic Procedures of the Chest, Extremities, Skull, Skeletal System, Respiratory System, Obstetric, Pelvis, Spinal Column And Gastro Intestinal Track And For Special Investigation Including Barium, IVP And HSG Examinations.


Sailent Features :-

 
 
 
  • Radiographic Output: 24 KW (PX-300) and 40KW (PX-500) as per IS7620 (Part 2)
  • Tube Filament Current Stabilisation Through Constant Voltage Transformer (CVT)
  • Full Wave Rectification
  • Digital Display of KV, mA and mAS
  • Overload Protection (Blockage Of Exposure)
  • Shortest Exposure Time 10 ms
  • Supports Serialography upto 6fpts
  • kv, mA and S valves Set With Step Switches
  • Floroscopy Time Totaliser
  • 5 Led Indicators for Line, O\ L ,Exposure, Ready and Bucky In,
  • Voltage Stabilisation (Compensation) Through Course and Fine Voltage Compensation Switches.
  • Volt Meter and mA meter for Viewing Input Line Voltage and mA during Exposure.
 
     
  Added   Applications :-  
     
  Providing You Complete Solutions For Every Application Wide Range Of Optional High Quality Accessories Are Also Available.All With The Guarantee Of Quality And Enduring Value That Only" PROFESSIONAL" Can Provide.


 
 
  • Radiography and Radiography \ Fluoroscopy Floor-To- Ceiling Column Stands
  • Horizontal and Manual Multi Position Bucky Examination Tables
  • Motor Operated Table Spot Film Device
  • Image Intensifier Television Option
  • Option of Single or Double Tube Operation
 
  Technical Data For PX – 300 \ 500 :-  
     
 

Specification

PX-300

PX-500

Output Power
(As Per IS : 7620)
24KW 40 KW
Radiography
Parameters
kV Range 40-100  kV (2  kV  Each  Step)
mA Range  Small Focus (50 ,100)
                     Large Focus (200 , 300)
kV  Range  40-100  kV  (2  kV  EACH  STEP)
mA Range   Small Focus (50 ,100 ,200)
                       Large Focus (300 , 400 ,500)
Exposure Time   0.01 to 5 Sec.    0.01 to 5 Sec.
Fluoro Scopy
Parameters
kV RANGE 40 TO 90 kV in Steps mA RANGE 0 -3 mA
Exposure Release From Control Console With Two Step Switch  or From Examination Unit
Fluorscopy Release From Control Console or From Examination   Unit
Mainsvoltage  Compensation Provided (354 -447 Volts)
Filament  Stabilisation Provided
Over Load Protection Provided , Blocking For Any Hazardous Combination
Tube Head THX-1  BEL\Imported Rotating Anode     Dual FOCUS 20\40 kW
                                                                                       OR
                                                                    Dual Focus 30\50 kW (Optional)           
HV Cable Pair of HV cable
Power Supply 3 Phase 440V+10%  - 15%  50Hz 63A
Main Resistance < . 36 OHM
 
     
( Powered by Kulhari Group )